Validity and reliability of a novel subtalar joint axis of rotation locator measurement device

Context

Inclination of the subtalar joint (STJ) in the transverse and sagittal planes may be highly associated with ankle sprain mechanisms. However, the validity and reliability of measuring inclination of the STJ axis of rotation (AoR) is not well established.

Objective

The purposes of this study were to: 1) examine the validity of a custom made instrument (locator) to measure the STJ AoR on the basis of the STJ inclination measured by X-ray, 2) to measure the intra-tester reliability of the locator.

Design

Cross sectional study.

Setting

Biomechanics laboratory.

Participants

Twenty nine healthy male (age: 22.89±9.11 yrs; weight: 77.68±18.32 kg; height: 176.16±14.16 cm) and Nine health female (age: 25±8 yrs; weight: 54.42±8.42 kg; height: 164.33±7.67 cm) subjects were recruited for this study.

Intervention

No Intervention.

Main outcome measures

Variables that were measured in this study were as follows: 1) Inclination of STJ AoR in the sagittal plane measured by radiographic images (Median MDXP-40 Inc, Korea) of the foot in the sagittal plane. In order to collect radiological images of the foot, subjects stood with a tandem position and the STJ was placed in neutral position. Sagittal plane inclination of the STJ AoR were further analyzed using ViewRex (TechHeim, Korea) per McClay‘s method [1]; 2) Inclination of the STJ AoR in the sagittal plane was measured by the locator; 3) Inclination of the STJ AoR in the transverse plane was measured by the locator. The anterior and posterior exit point were determined per Kirby’s method [2]. Once the locator was aligned along two points, (anterior medial and posterior lateral exit point) a Digital Mini Protractor (WWC-TE Bevel Box, USA) was used to measure inclination angle. Pearson correlation was used to analyze the relationship of validity between radiographs and the locator measuring the STJ sagittal plane inclination. Intra Correlation Coefficient (ICC) was used to analyze day-to-day reliability of the locator.

Results

Table 1 Intra-test reliability about STJ sagittal plane.

Conclusion

The locator may be used in the clinical setting since validity verified by correlation was high and the intra-test correlation coefficient was large indicating consistent measurements. Along with the locator measurement, it is suggested that further study including motion analysis may provide more information regarding the relationship between inclination of STJ AoR and movement at the STJ.

References

  1. McClay I, Bray J: The Subtalar Angle: A Proposed Measure of Rearfoot Structure. Foot Ankle Intl. 1996, 17 (8): 495-502.

    Article  Google Scholar 

  2. Spooner SK, Kirby KA: The Subtalar Joint Axis Locator. J Am Podiatr Med Assoc. 2006, 96 (3): 212-219.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to SY Lee.

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

The Creative Commons Public Domain Dedication waiver (https://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Cite this article

Kim, B., Lee, S., Lee, H. et al. Validity and reliability of a novel subtalar joint axis of rotation locator measurement device. J Foot Ankle Res 7 (Suppl 1), A41 (2014). https://doi.org/10.1186/1757-1146-7-S1-A41

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1757-1146-7-S1-A41

Keywords