Table 1 Results from the on-the-bench and on-site assessment, and with respect to some clinically relevant parameters.

From: Pressure measurement devices: from technical assessment to clinical performance

PMD under test ISS Full technical assessment ISS On-site partial assessment “gait” assessment: Peak pressure (kPa) “gait” assessment: Mean pressure (kPa)) “gait” assessment: Integral (kPa*s)[2]
a not performed error >10% at 250kPa 100 (4)** 80 (2)** 39 (2)**
b not performed error < 5% at 250kPa 266 (12)* 191 (8)* 85 (9)*
C accuracy error < 5% up to 1200kPa error < 5% at 250kPa 744 (137) 367 (17) 152 (23)
  1. * statistically different from PMDc corresponding data (p<0.05, also verified with respect to the ± 5% maximum error); ** statistically different from PMDb and PMDc corresponding data (p<0.05, also verified with respect to the ± 5% maximum error)