Skip to main content

Table 4 Diagnostic accuracy of Vibrasense and monofilament tests compared to NCS

From: Assessment of the diagnostic accuracy of Vibrasense compared to a biothesiometer and nerve conduction study for screening diabetic peripheral neuropathy

Tests comparison

NCS diagnosis

Normal (N = 33)

Abnormal (N = 28)

Test diagnosis

Positive

Negative

Positive

Negative

Vibrasense

7

26

23

5

10-g monofilament

3

30

10

18

Diagnostic accuracy

Parameter

Sensitivity

Specificity

PPV

NPV

Vibrasense

82.14%

78.79%

76.67%

83.87%

10-g monofilament

35.7%

90.90%

76.92%