Table 4 Reliability and Measurement error results

From: Elastography in the assessment of the Achilles tendon: a systematic review of measurement properties

Outcome Measurement Instrument

Author and year

Reliability

Measurement Error

Study population

Design

Cosmin Rate

Statistical Results

Measurement property ratings based on statistical results

Cosmin Rate

Statistical Results

Measurement property ratings based on statistical results

Strain elastography

Drakonaki et al., 2009 [33]

25 bilateral

Inter-rater same day

Adequate

ICC T = 0.41 L = 0.51

-

Adequate

CV T = 30% L = 29.6%

?

Intra rater same day

ICC T = 0.41,0.45 L = 0.78,0.66

 + 

CV T = 39% L = 30.50%

?

Ooi et al., 2015 [27]

10 bilateral

Inter rater different days

Very good

ICC = 0.79

 + 

NA

NA

NA

Intra rater different days

Very good

ICC = 0.87

 + 

Yamamoto et al., 2016 [34]

25 bilateral

Inter rater same day

Doubtful

Spearman = 0.61

NA

Adequate

SEM = 0.06,0.07

?

Intra rater same day

Adequate

ICC = 0.93 ICC = 0.87

 + 

Schneebeli et al., 2016 [35]

24 bilateral

Intra rater same day

Adequate

ICC relaxed = 0.87 ICC contracted = 0.94

 + 

NA

NA

NA

Payne et al., 2017 [32]

8 right

Intra rater same day

Intra rater different days

Adequate

ICC T = 0.00–0.11, L = 0.01–0.11 both foot position

-

Adequate

CV L = 80.8%-111.5% T = 53.6—112.4%

?

Shear wave elastography—velocity

Aubry et al., 2013 [38]

30 bilateral

Inter-rater same day

Doubtful

ICC L = -0.011–0.46, T-0.062–0.29

NA

NA

NA

NA

DeWall et al., 2014 [39]

10 feet side not reported

Inter-rater

Doubtful

CV = 0.156

NA

NA

NA

NA

Fu et al., 2016 [40]

326 bilateral

Inter-rater same day

Adequate

ICC L = 0.923, T = 0.870

 + 

NA

NA

NA

Payne et al., 2018 [47, 48]

3 dominant foot

Inter-rater same day

Very good

ICC T-0.70 L-0.80

 + 

Very good

SEM T = 0.19–0.26 L = 0.19–0.27 CV T = 3.9–6.3% L = 2.9–5.2%

?

14 dominant foot

Intra rater same day

Intra rater different days

ICC T = 0.62–0.85, L = 0.45–0.71

-

Coombes et al., 2018 [30]

6 bilateral

Inter session different days

Very good

ICC = 0.71

 + 

NA

NA

NA

Götschi et al., 2021 [46]

10 right

Inter rater different days

Adequate

ICC = 0.455

-

Adequate

SEM = 1.043 m/s

?

Inter session different days

Adequate

ICC = 0.591

-

Adequate

SEM = 1.068 m/s

?

Shear wave elastography—modulus

Helfenstein-Didier et al., 2016 [41]

7 right

Intra session same day

Adequate

ICC s = 1.0 kPa c = 0.42–0.85 MPa

-

Adequate

SEM s = 1.37–2.14 kPa, c = 3.5–10.69 MPa

?

Lima et al., 2017 [42]

24 bilateral

Intra session different days

Doubtful

ICC = 0.82–0.93

NA

Doubtful

CV = 23–25%

NA

Inter sessions different days

ICC = 0.42–0.60 to isometric contraction

Zhou et al., 2019 [43]

20 dominant foot

Inter rater different days

Very good

ICC = 0.76–0.94

 + 

Very good

SEM = 14.38–15.78 kPa

 + 

Intra rater different days

ICC = 0.77–0.93

 + 

SEM = 11.87–21.75 kPa

 + 

cSWE

Suydam et al., 2015 [44]

29 bilateral

Intra rater same day

Adequate

ICC s = 0.875, v = 0.876

 + 

Adequate

SEM s = 3.8 kPa v = 6.8Pas

 + 

Corrigan et al., 2019 [45]

20 either right or left

Intra rater same day

Intra rater different days

Adequate

ICC s = 0.697, v = 0.856, d = 0.855

 + 

Adequate

SEM s = 8.28 kPa, v = 4.79 Pas, d = 46.72 kPa

?

3D SWE

Götschi et al., 2021 [46]

10 right

Inter rater different days

Adequate

ICC = 0.436

-

Adequate

SEM = 0.553 m/s

?

Inter sessions different days

ICC = 0.591

-

SEM = 0.505

?

  1. NA Not applicable, ICC Intraclass correlation coefficient, CV Coefficient of variance, SD Standard deviation, SEM Standard error of measure, L Longitudinal, T Transverse, c Compression modulus, s Shear modulus, v Viscosity modulus, d Dynamic modulus