From: Validation and reliability testing of a new, fully integrated gait analysis insole

Fig. 3

Validation results for peak force (a) and stance time (b) measurements. The intraclass correlation coefficients for single, as well as average measures are shown. The corresponding Bland-Altman plot is shown beneath each table. Bar graphs with standard deviation for resultant force (c) and step time (d) are shown. No significant differences between both systems for resultant force (difference between means: 36.3 ± 27.19 N; p = 0.19), as well as stance time (difference between means: 0.027 ± 0.028 s; p = 0.36) were seen. Data from 360 steps over the force plate are shown