Table 1 Changes in ROM after the sensor insertion for all specimens and bone combinations

From: Insertion of a pressure sensing arrayminimally affects hindfoot bone kinematics

Plane   tib-tal tal-calc tal-nav calc-cub nav-cub
SAGITTAL Specimen 1 -1.3 -0.5 -0.8 0.4 -0.1
  Specimen 2 -4.2 -0.7 0.7 0.1 -0.4
  Specimen 3 3.8 -9.5 * -10.6 * -0.3 0
  Specimen 4 0 -7.8 * -0.7 -14.5 * -0.9
  Specimen 5 6.7 * 2.1 5.7 * 1.2 -1.5 *
  Specimen 6 -7.9 -2.4 4.3 0.1 -0.1
  Specimen 7 -2.3 3.2 4.9 1.1 -0.2
  All Specimens -0.9 -0.7 1 -0.1 -0.3
  StD 6.7 6.4 4.9 5.6 1.1
CORONAL Specimen 1 0.1 -0.4 -0.6 0.3 -0.5
  Specimen 2 0.4 -0.7 -2.3 2 -1
  Specimen 3 -14.1 * -10.1 * 2 1 1.4
  Specimen 4 0.9 -0.4 -1.9 -12.3 * -1.4
  Specimen 5 3.7 1.3 -2.8 0.4 0.9
  Specimen 6 8.5 * -2 -4.9 3.2 3.8 *
  Specimen 7 -1.2 -1.5 -3.4 1.7 -1.9 *
  All specimens -0.9 -1.2 -1.6 0.1 -0.6
  StD 4.4 4.1 5.4 4.4 1.7
TRANSVERSE Specimen 1 -0.2 0.6 0 0.2 -0.8
  Specimen 2 2.5 1.8 -3.5 -0.3 0.7
  Specimen 3 -16.4 * -15.9 * -34.3 * 0.9 -1.4
  Specimen 4 -0.3 -0.8 -4.2 -8.3 * 0.8
  Specimen 5 3.5 1.4 -4.7 3.8 * -0.1
  Specimen 6 1.6 3.7 -1.5 0.2 -0.4
  Specimen 7 -2 -1.9 7.7 2.8 * 2 *
  All specimens 0.6 -0.1 -3.2 -0.6 -0.2
  StD 6.3 5 10.8 2.6 1.9
  1. Statistical significant differences within the standard deviation rate are reported with a star (*, p<0.05)