Skip to main content

Table 1 Exposure parameters of the CBCT and MSCT scanners and conventional X-ray devices

From: Effective radiation dose of a MSCT, two CBCT and one conventional radiography device in the ankle region

 

Siemens

NewTom 5G

Planmed

Shimadzu

 

Somatom

CBCT

Planmed

FH-21 HR

 

Definition AS+

Hi Res

Standard

Standard

Verity

AP + LAT

 

MSCT

“12 × 8”

“15 × 12”

“18 × 16”

CBCT

Radiography

Potential (kV)

120

110

110

110

90

57

Tube current (mA)‡

54

11

1,5

0,6

8

-

Exposure time (s)

1.0

5,4

3,6

3,6

6

-

Q (mAs)

54

59

5,3

2,3

48

16

Slice thickness (mm)

0.6

-

-

-

-

-

Pitch (mm)

0.5

-

-

-

-

-

Voxel H (mm)

-

0.15

0.3

0.3

0.2

-

Voxel L (mm)

-

0.15

0.3

0.3

0.2

-

Voxel W (mm)

-

0.15

0.3

0.3

0.2

-

Scan angle

360°

360°

360°

360°

210°

-

Frame number

-

360

360

360

300

-

Scan FOV (mm)†

130 × 160

120 × 80

150 × 120

180 × 160

130 × 160

130 × 160

  1. †For MSCTs and CBCT: diameter × length, for conventional X-ray width × height.
  2. ‡For NewTom 5G CBCT: mA is automatically adjusted using the “SafeBeam™” acquisition option.