Skip to main content

Table 2 Within session reliability

From: Reliability of measuring abductor hallucis muscle parameters using two different diagnostic ultrasound machines

 

ICC [3.1]

95% CI

SEM

SRD

Phillips HD11

    

Dorso-plantar thickness (mm)

0.97

(0.95 to 0.99)

0.12

0.34

Medio-lateral width (mm)

0.96

(0.95 to 0.98)

0.87

2.47

Cross-sectional area (mm2)

0.98

(0.96 to 0.98)

9.36

26.59

Chison 8300

    

Dorso-planatr thickness (mm)

0.99

(0.99 to 0.99)

0.09

0.26

Medio-lateral length (mm)

0.95

(0.92 to 0.97)

1.06

3.01

Cross-sectional area (mm2)

0.99

(0.98 to 0.99)

6.73

19.12