Skip to main content

Table 4 Equivalent dose contributions and effective doses (μSv) attained in the ankle region using MSCT, CBCT and radiography devices

From: Effective radiation dose of a MSCT, two CBCT and one conventional radiography device in the ankle region

 

Siemens

NewTom 5G

 

Shimadzu

 

Sensation

CBCT

Planmed

FH-21 HR

 

Open

Hi Res

Standard

Standard

Verity

X-ray

 

MSCT

“12 × 8”

“15 × 12”

“18 × 16”

CBCT

AP+LAT

Bone marrow

9.5

3.9

1.8

0.8

2.7

0.7

Bone surface

7.2

7.3

1.2

0.6

1.9

0.3

Skin

3.1

2.2

0.7

0.3

0.9

0.3

Remainder

      

Lymphatic nodes

0.9

0.8

0.2

0.1

0.3

0.1

Muscles

0.6

0.1

0.1

0.1

0.1

0.1

Effective dose

21.4

14.3

4.0

1.9

6.0

1.5