Reliability of hand held Doppler use in podiatrists
Journal of Foot and Ankle Research volume 8, Article number: O7 (2015)
Background
Hand held Doppler examination is a frequently used non-invasive vascular assessment utilised by podiatrists. Despite this, Doppler's reliability has not been thoroughly investigated. Given the importance of Doppler in completing a vascular assessment of the lower limb, it is essential to determine the reliability of the testing method in practicing podiatrists.
Methods
This was a multi-centre inter-tester and intra-tester reliability study. Four podiatrists participated in this study, two public and two private practitioners. Three aspects of Doppler use were examined; the clinical Doppler technique, the evaluation of Doppler audio sounds and qualitative evaluation of Doppler waveforms. Participants meeting current guidelines for vascular screening attended two testing sessions, one week apart at either the private practice (N=32), or the public practice (N=31). Two podiatrists (either public or private) assessed their Doppler waveforms, and rated them as mono-phasic or multi-phasic. Podiatrists were also then required to analyse 30 waveform tracings chosen at random by the researchers, and 30 audio recordings of Doppler sounds recorded by the researchers. Cohen's Kappa (K) statistics were used to calculate inter and intra tester reliability using SPSS version 19.
Results
Evaluation of waveform tracings demonstrated highest reliability, with inter-tester reliability ranging from K 0.77 to 0.90, and intra-tester reliability from K 0.81 to 1.00. The public podiatrists showed higher reliability in audio examination (inter-tester reliability K 0.61, intra-tester reliability K 1.00) compared to the private podiatrists (inter-tester reliability K 0.31, intra-tester reliability K 0.53). Clinical use of Doppler demonstrated the lowest reliability for both pairs of clinicians (inter-tester reliability K 0.20 to 0.24, and intra-tester reliability K 0.27 to 0.42).
Conclusions
There is a need for ongoing education for podiatrists using Doppler in clinical practice, as the reliability of the clinical use of the Doppler was low. This indicates that technique could be an issue. There is also a need to evaluate if hand held Doppler equipment is suitable for use in the podiatry patient cohort.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Craike, P., Schlachter, B., Tehan, P. et al. Reliability of hand held Doppler use in podiatrists. J Foot Ankle Res 8 (Suppl 2), O7 (2015). https://doi.org/10.1186/1757-1146-8-S2-O7
Published:
DOI: https://doi.org/10.1186/1757-1146-8-S2-O7